Table 1. Uncertainty budget for amplitude

Sources of uncertainty Standard uncertainty for amplitude (mV) Type Probability distribution
u(Ap) 0.34 A Normal
u(Ap. repeatability) 0.34
u(Ap. resolution) 0.003
u(Ap. resolution) 0.003 B Rectangular
u(CA) 0.2 B Normal
uc(Ap) 0.44 Assumed as rectangular
U(Ap) 0.88 Assumed as rectangular